Designing & Reporting Spectral Workflows¶
Purpose: Design scientifically sound workflows and report results reproducibly.
Audience: Spectroscopy researchers planning new analyses; data scientists building pipelines.
Time: 20–30 minutes to design; reference while building workflows.
Prerequisites: Understand FoodSpec data model (spectra, metadata); basic experimental design.
This chapter explains what a workflow means in FoodSpec and how to design, execute, and report it. A workflow connects the scientific question, instrument, data acquisition, preprocessing, feature extraction, statistics/ML, interpretation, and reporting.
What is a workflow in FoodSpec?¶
flowchart LR
subgraph Data
A[Scientific question]
B[Experimental design & instrument]
C[Data acquisition & organization]
end
subgraph Preprocess
D[Baseline / smoothing / norm / crop]
E[Features: peaks, ratios, PCA/PLS]
end
subgraph Model_Stats
F[Stats & ML (tests, classifiers, regressors, mixtures)]
end
subgraph Report
G[Metrics + plots + report.md + run_metadata]
end
A --> B --> C --> D --> E --> F --> G
- It is a pipeline from question → instrument → data → preprocessing → features → analysis → interpretation → reporting.
- It aligns scientific goals with acquisition conditions, preprocessing choices, analysis methods, and clear outputs.
- Model and metric choices are covered in ML & DL models and Metrics & evaluation.
- For issues at any stage (baseline, SNR, imbalance, overfitting), see Common problems & solutions.
Quick Example: Oil Authentication Workflow¶
from foodspec.io import load_csv
from foodspec.preprocess import baseline_als, normalize_snv
from foodspec.ml import train_classifier
# 1. Load data
ds = load_csv("oils.csv", wavenumber_col="wavenumber")
print(f"Loaded {len(ds)} spectra with {len(ds.wavenumbers)} wavenumbers")
# 2. Preprocess
ds = baseline_als(ds, lam=1e6)
ds = normalize_snv(ds)
# 3. Train classifier
model, metrics = train_classifier(
ds,
label_column="oil_type",
method="random_forest",
cv_strategy="stratified_kfold"
)
print(f"Accuracy: {metrics['accuracy']:.2%}")
How to design a workflow (step-by-step)¶
- Clarify the question. Examples: “Can we distinguish oil types?”, “How does heating affect quality?”, “Is this batch within spec?”
- Identify response variables and factors. Class labels (oil_type), continuous metrics (peroxide value, mixture fraction), time/temperature, batches.
- Consider sample types/matrices. Pure oils vs oils in chips; liquid vs powder; expected heterogeneity.
- Choose instrument and acquisition conditions. Spectral range, resolution, SNR, laser power/integration time; ATR vs transmission; calibration routines.
- Plan preprocessing and features. Baseline/smoothing/normalization; peaks/ratios, PCA/PC scores; see preprocessing chapters.
- Plan statistics/ML. Are you comparing groups (ANOVA/t-test), predicting labels (classification), estimating fractions (regression/NNLS), detecting outliers (QC)?
- Plan visualization and reporting. What plots/figures and metrics will demonstrate your answer?
Questions to ask at each stage¶
- Scientific goal: What decision/action will this analysis support? What is “good enough” performance?
- Instrument & acquisition: What range/resolution is needed? What SNR? Are calibration and baselines stable? Any fluorescence, saturation, or drift?
- Data structure: How many samples per group? Are groups balanced? Any batch effects or confounders? Randomization?
- Analysis: Comparing groups, predicting labels, quantifying mixtures, detecting novelty? Expect linear/monotonic trends or complex patterns?
- Reporting: What should a reviewer reconstruct? Which methods, parameters, metrics, and figures must be included?
Tools & methods in FoodSpec (questions → approaches)¶
| Question | Typical approach | FoodSpec modules/functions | Example workflow |
|---|---|---|---|
| Differentiate classes (e.g., oil types) | Classification (RF/SVM/PLS-DA), ANOVA on key ratios | apps.oils, preprocess.*, features.*, stats.run_anova |
Oil authentication |
| Track degradation over time | Ratios vs time, correlation/linear regression, ANOVA across stages | apps.heating, stats.compute_correlations, stats.run_anova |
Heating quality monitoring |
| Estimate mixture proportions | NNLS/PLS regression, correlation of predicted vs true | chemometrics.mixture, stats.compute_correlations |
Mixture analysis |
| Screen batches for anomalies | One-class SVM/IsolationForest, t-tests on key metrics | apps.qc, stats.run_ttest |
Batch QC |
| Assess associations | Pearson/Spearman, cross-correlation for sequences | stats.compute_correlations, compute_cross_correlation |
Heating/time series |
| Visualize results | Confusion matrices, PCA, correlation heatmaps, calibration plots | viz.confusion_matrix, viz.plot_pca_scores, viz.plot_correlation_heatmap, viz.plot_regression_calibration |
Across workflows |
When assumptions are doubtful or samples are small/skewed, use nonparametric tests (run_mannwhitney_u, run_kruskal_wallis, run_wilcoxon_signed_rank, run_friedman_test) and robustness checks (bootstrap_metric, permutation_test_metric) from foodspec.stats.
For instrument/file formats and ingestion routes, see the Instrument & file formats guide.
See also:
- Preprocessing: Baseline, Normalization, Feature extraction
- ML: Classification & regression, Mixture models
- Stats: Hypothesis testing, ANOVA/MANOVA
- Protocols: Reproducibility checklist
- API: foodspec.preprocess.baseline.ALSBaseline, foodspec.stats.run_anova, foodspec.apps.oils.run_oil_authentication_workflow
Instrument specs, calibration, limitations → data¶
- Range & resolution: Determines which bands are captured/resolved. Choose ranges that cover key functional groups.
- SNR & dynamic range: Low SNR → more smoothing/replicates; saturation/bleed → adjust integration/laser power.
- Laser power & integration time: Affect intensity and fluorescence; document settings.
- Optics (ATR vs transmission): Influences depth of penetration and baseline shape.
- Calibration: Wavenumber calibration (e.g., silicon peak), intensity calibration if needed. Drift appears as shifts/baseline changes.
- Artefacts: Baseline drift, fluorescence, noise, cosmic rays. These motivate baseline correction, smoothing, normalization, and spike removal.

Reporting checklist (workflow level)¶
- Instrument: Model, spectral range/resolution, laser power/integration, calibration routine.
- Samples: Type/matrix, preparation, storage, number of replicates per group.
- Acquisition: Order/randomization, accumulations, environmental conditions.
- Preprocessing: Steps and parameters (baseline λ/p, smoothing window/poly, normalization choice, cropping).
- Features: Peak/ratio definitions, band ranges, PCs used.
- Analysis: Models, hyperparameters, validation strategy; statistical tests (alpha, effect sizes).
- Metrics: Accuracy/F1/ROC for classification; RMSE/MAE/R² for regression; test statistics/p-values/effect sizes for stats.
- Figures: Spectrum overlays, mean±CI spectra, PCA scores/loadings, confusion matrices, boxplots/violin plots, correlation plots, residuals.
- Data/code: Where to find raw/processed data, configs, run_metadata.json, reports.
Plots & visualizations: what, why, and when¶
| Plot type | Main use | Example question | FoodSpec context |
|---|---|---|---|
| Raw/mean spectra | Inspect quality, baseline, noise | Are baselines drifting? | Preprocessing QC |
| PCA scores/loadings | Explore structure, drivers | Do oils cluster? Which bands matter? | Oil auth |
| Confusion matrix | Classification performance | Where does the classifier err? | Oil auth/QC |
| Box/violin plots | Group comparisons | Do ratios differ by type/stage? | Stats on features |
| Correlation scatter/heatmap | Association strength | Do ratios correlate with time/quality? | Heating/mixtures |
| Residual plots | Model diagnostics | Are residuals patterned? | Regression/mixtures |
| Time/temperature trends | Process monitoring | How do markers change with heating? | Heating workflow |
Expected details: axes labels (units), legends, group labels, sample sizes, confidence intervals/shading where relevant, consistent wavenumber units (cm⁻¹).
Example mini-workflows¶
Oil authentication (short narrative)¶
- Question: distinguish oil types/adulteration.
- Design: multiple oil classes, balanced reps; randomize acquisition; Raman fingerprint range.
- Acquisition: set laser/integration to avoid fluorescence/saturation; calibrate wavenumber.
- Preprocess: ALS baseline, Savitzky–Golay smoothing, L2 norm, crop.
- Features: peaks 1655/1742/1450; ratios.
- Analysis: RF classifier + stratified CV; ANOVA/Tukey on ratios; PCA for exploration.
- Plots: PCA scores/loadings, confusion matrix, boxplots of ratios.
- Reporting: metrics ± variability, test stats/effect sizes, preprocessing/model params.
Heating quality monitoring (short narrative)¶
- Question: track degradation markers vs time/temperature.
- Design: time points/stages with replicates; control batches; randomize order if possible.
- Acquisition: consistent settings; monitor drift.
- Preprocess: baseline, smoothing, normalization, crop.
- Features: unsaturation ratio (1655/1742) vs time; optional other bands.
- Analysis: correlation/linear regression of ratio vs time; ANOVA if discretized stages; trend models.
- Plots: ratio vs time with fit, boxplots by stage, optional residuals.
- Reporting: slope/p-value, R², CIs; preprocessing/model details.
Summary¶
- A workflow is a coherent pipeline from question to report; design drives instrument choice, preprocessing, analysis, and visualization.
- Ask the right questions at each stage; use FoodSpec tools (preprocessing, features, ML, stats) accordingly.
- Report instruments, preprocessing, features, analysis methods, metrics, and figures transparently.
When Results Cannot Be Trusted¶
⚠️ Red flags for workflow validity:
- Instrument or preprocessing not documented ("we used standard Raman; no special preprocessing")
- Results depend critically on instrument settings and preprocessing choices
- Undisclosed choices prevent reproduction or cross-instrument validation
-
Fix: Document laser wavelength, power, integration time, baseline correction (parameters), smoothing, normalization, peak alignment
-
No reference sample or control in workflow (no validation against known composition)
- Unknown whether conclusions apply to intended samples or are artifacts
- Control validates method, detector, and preprocessing
-
Fix: Include positive control (known authentic/adulterant), negative control (pure reference), and blank (empty cuvette)
-
Workflow tested only on same instrument/day/operator
- Results don't generalize to other labs or future analyses
- Batch effects (drift, calibration) confound biological effects
-
Fix: Validate on data from different instruments, days, or operators; use batch-aware CV; test generalization
-
Preprocessing parameters tuned on same data used for model evaluation
- Overfitting in preprocessing (hyperparameter tuning) inflates metrics
- Parameters optimized on test data don't generalize
-
Fix: Freeze preprocessing parameters before data splitting; include preprocessing in CV loop
-
Results meeting workflow goals without investigating outliers or failures
- Some samples may fail preprocessing (e.g., fluorescing oils, broken cuvette)
- Ignoring failures overestimates success rate
-
Fix: Document failure modes; report success rate; investigate failed samples
-
No comparison to baseline or previous method (reporting new workflow accuracy with no context)
- Accuracy/precision depend on problem difficulty; comparison shows improvement
- New workflow may be worse than old if not benchmarked
-
Fix: Compare metrics to baseline (previous method, random guess, published results); report relative improvement
-
Temporal structure ignored (time-series workflow: training on future data, testing on past)
- Leakage through time produces overoptimistic metrics
- For trending workflows (heating quality), time-aware CV needed
-
Fix: Use temporal CV (train on past, test on future); no forward-looking leakage
-
Interpretation relies only on model metrics without visual/chemical validation
- Metrics alone don't confirm biological/chemical validity
- Workflow producing high accuracy on noise still worthless
- Fix: Visualize outputs (spectra, residuals, predictions); validate top features chemically; involve domain experts